HT instruments HT701 User Manual

Page 22

Advertising
background image

HT701

EN - 21

Resistance

Range Resolution Accuracy

Max voltage

open circuit

Overload protection

1000.0

 0.1

(0.5%rdg+2dgt)

Approx. 0.25V

1000VDC/ACrms

10.000k

 0.001k

100.00k

 0.01k

1000.0k

 0.1k

10.000M

 0.001M

40.00M

 0.01M

Continuity test

Range Accuracy Buzzer

Open circuit voltage Overload protection

400.0

(0.5%rdg+2dgt) <30

Approx. 1.2V

1000VDC/ACrms

Diode test

Range

Accuracy

Test current Open circuit voltage Overload protection

2.000V

(0.5%rdg+2dgt)

0.6mA 2.5V

1000VDC/ACrms

Frequency of AC Voltage and AC Current

Range Resolution Accuracy

Minimum duration

of pulse

Overload protection

100.00Hz 0.01Hz

(0.1%rdg+5dgt)

10µs

1000VDC/ACrms

max 440mA

1000.0Hz 0.1Hz

10.000kHz 0.001kHz
100.00kHz 0.01kHz

Signal sensitivity for frequency measure

Function Range

Sensitivity (sinusoidal waveform)

10Hz ÷ 10kHz

10kHz ÷ 100kHz

AC mV

100.00mV 15.00mV
1000.0mV 150.0mV

AC V

10.000V 1.500V
100.00V 3V

-

1000.0V 30V

-

AC mA

100.00mA 15.00mA

-

400.0mA 30mA

-

Capacitance

Range Resolution Accuracy Measuring

time

Overload

protection

10.000nF 0.001nF

(1.2%rdg+80dgt)

0.7s

1000VDC/ACrms

100.00nF 0.01nF

(1.2%rdg+20dgt)

1000.0nF 0.1nF

(1.2%rdg+2dgt)

10.000

F 0.001F

100.00

F 0.01F

1000.0

F 0.1F

3.75s

10.000mF 0.001mF

(1.2%rdg+20dgt)

7.5s

40.00mF 0.01mF

(1.2%rdg+80dgt)

Advertising