HT instruments HT701 User Manual

Page 23

Advertising
background image

HT701

EN - 22

Temperature with K probe

Range Resolution Accuracy

Overload

protection

-200.0°C ÷ 0.0°C

0.1°C

(1.0%rdg+2°C)

1000VDC/ACrms

0.0°C ÷ 1200.0°C

(1.0%rdg+1°C)

-328.0°F ÷ 32.0°F

0.1°F

(1.0%rdg+36°F)

32.0°F ÷ 2192.0°F

(1.0%rdg+18°F)


Insulation resistance

Test voltage

Measuring range

Accuracy Overload

protection

50V DC

2.000M

(1.5%rdg+5dgt)

600VDC/ACrms

20.00M

55.0M

100V DC

2.000M

20.00M

110.0M

250V DC

2.000M

20.00M

200.0M

275M

500V DC

2.000M

20.00M

200.0M

550M

1000V DC

2.000M

20.00M

200.0M

2000M

22.0G

(10%rdg+3dgt)

Test voltage accuracy: +20%rdg, -0%rdg

Short-circuit current: 1mA

Minimum resistance (@ nominal current 1mA): 50k

 (50V), 100k (100V), 250k (250V), 500k (500V), 1M (1000V)

Discharge time of measured object: <1s (C

 1µF)

Maximum capacitive load: 1µF

Voltage detection on circuit: test inhibited for voltages

30V AC/DC on inputs