HT instruments HT701 User Manual

Page 24

Advertising
background image

HT701

EN - 23

6.1.1. Electrical

characteristics

Conversion: TRMS
Sampling frequency:

3 times per second

Temperature coefficient:

0.15x(accuracy) /°C, <18°C or >28°C

NMRR Normal Mode Rejection Ratio:

> 50dB for DC quantities and 50/60Hz

CMRR Common Mode Rejection Ratio:

>100dB from DC up to 60Hz (DCV)

> 60dB from DC, up to 60Hz (ACV)


6.1.2. Considered

standards

Safety: IEC/EN

61010-1

IEC/EN61557-1,

IEC/EN61557-2

Insulation:

double insulation

Pollution level:

2

Overvoltage category:

CAT IV 600V, CAT III 1000V

Max operating altitude:

2000m (6561ft)



6.1.3. General

characteristics

Mechanical characteristics
Size (with shell):

207 (L) x 95 (W) x 52 (H) mm (8 x 4 x 2 inch)

Weight (batteries included):

630g (22 ounces)


Power supply
Battery type:

4 x 1.5V alkaline type AA IEC LR6

Low battery indication:

symbol “ ” with battery voltage < approx. 4.8V

Battery life:

approx. 80 hours
600 insulation test with new batteries at room
temperature (1M

 @ 1kV, duty cycle of 5s on

and 25s off)

Auto Power Off

after 20 minutes of idleness (disabled)


Memory
Characteristics:

max. 100 locations for each function


Display
Characteristics:

5 LCD, 10000 points, decimal sign, analogue
bargraph and autobacklight

Over range indication:

OL” or “-OL


6.2. ENVIRONMENT
6.2.1. Environmental conditions for use
Reference temperature:

23° ± 5°C (73 ± 41°F)

Operating temperature:

0° ÷ 50°C (32° ÷ 122°F)

Allowable relative humidity:

<80%RH

Storage temperature:

-20° ÷ 60°C (-4 ÷ 140°F) (batteries not fitted)

Storage humidity:

<80%RH (batteries not fitted)


This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC

(LVD) and of EMC Directive 2004/108/EC