HT instruments M70 User Manual

Page 12

Advertising
background image

M70

EN - 11

4.5.1. "ZERO"

mode

1. Any addition or replacement of cables, extensions and croco clips nullify the previous

calibration and make necessary a new calibration before performing further
measurements. Therefore the instrument must be calibrated in the same conditions at
which it will operate during measurements

2. Short-circuit the cable ends with each other as shown in Fig. 7 making sure that the

metallic parts of test probes and crocodiles are in good touch

Fig. 7: Connection of the instrument’s terminals during calibration procedure

3.

GO

Press GO to perform the measurement

4.

MODE

PEAK

Press and hold ZERO key for 2s. The instrument resets the resistance of the
cables, the symbol ZERO is displayed

5. The measured value is stored by the instrument and used as offset (which means it’s

deducted from all continuity tests performed) until a new pressing of the ZERO key for
2s that voids the calibration

6. Each time the instrument is switched off or changing position of selector, the

calibrated value is lost

4.5.2. Anomalous

cases

which may occur during Lo

measurement

1. The full scale of the instrument is

19.99

. If the resistance value is higher

than this value, or in case of open or
interrupted probes, the instrument
displays the screen beside

2. If the voltage present at the terminals is

higher than 24V the instrument does not
perform the test. The screen beside is
displayed