Technical specifications – HT instruments M70 User Manual

Page 16

Advertising
background image

M70

EN - 15

6. TECHNICAL SPECIFICATIONS

6.1. CHARACTERISTICS
Accuracy is indicated as [%reading + (digit number*resolution)] at 23°C±5°, RH <70%.

DC Voltage

Range Resolution Accuracy

Input impedance

Overload protection

0.1

 600.0V

0.1mV

(0.5 rdg + 1 dgt)

3M

605V AC max RMS

AC Voltage

Range Resolution

Accuracy

(40 ÷ 500Hz)

Input impedance

Overload protection

0.1

 600.0V

0.1mV

(0.8 rdg + 4 dgt)

3M

605V AC max RMS

Max crest factor:

2

Resistance and Test Continuity

Range Resolution

Accuracy

Overload

protection

0.0

 199.9 0.1

(2.0% rdg + 3 dgt)

605V AC max RMS per 1 minute

The buzzer sounds while measured resistance is lower than 2

0.2A: Continuity test

Range Resolution

Accuracy

Overload

protection

0.00

 19.99 0.01

(2.0% rdg + 3 dgt)

605V max RMS

Test current: >200mA DC up to 5

 (measuring cables resistance included)

Test current: >10mA DC over 5

 (measuring cables resistance included)

Open circuit voltage: 4 < V

0

< 24V

M

: Insulation resistance

Test voltage

Range

Resolution

Accuracy

Overload protection

250V

0.001

 0.100M 0.001M

10dgt

605V max RMS

0.101

 3.999M 0.001M

(2.0%rdg+ 5dgt)

4.00

 39.99M 0.01M

40.0

 399.9M 0.1M

(5.0%rdg +5dgt)

400

 1000M 1M

500V

0.001

 0.250M 0.001M

15dgt

0.251

 3.999M 0.001M

(2.0%rdg +5dgt)

4.00

 39.99M 0.01M

40.0

 399.9M 0.1M

400

 2000M 1M

(5.0%rdg +5dgt)

1000V

0.001

 0.250M 0.001M

15dgt

0.251

 3.999M 0.001M

(2.0%rdg +5dgt)

4.00

 39.99M 0.01M

40.0

 399.9M 0.1M

400

 1000M 1M

(3.0%rdg +5dgt)

1000

 4000M 1M

(5.0%rdg +10dgt)

Autorange
Open circuit voltage: <1.3 x V

0

Accuracy of nominal voltage: –0% +10%
Short circuit current: <15mA
Nominal testing current: 1mA @ 1K

 x V (1mA @ 500K)

Advertising