Check Instrument Depth
At this point, if convenient, you can connect the Aqua

TROLL 200 to a PC, launch the software, and take a

reading. If the instrument is at the desired depth, secure it

in position as suggested below, or reposition as necessary.

If you set the software to “Remind me later” to set a Level

Reference, enter the level reference after installation when

prompted.

Secure the Cable
The RuggedCable includes a Kellems® grip near the

surface end. You can slide it along the cable to the desired

position by compressing it. When you pull on it, the grip

tightens and stops sliding. You may need to pull on both

ends of the Kellems grip to properly tighten it and keep it

from slipping.

Use the loop of the Kellems grip to anchor the cable to a

convenient stationary object. It works well with In-Situ’s

well dock installation ring. Simply insert the loop into the

locking clip on the well dock, and position the assembly

on the top of a well.

Installation Tips
4

Never let a probe fall freely down a well. The resulting

shock wave when it hits the water surface can damage

the strain gauge.

4

Check the level of water above the probe, then move

it and read again to be sure that the probe presents a

reasonable reading and shows change. It might not

be located where you think it is. For example, it could