Relay connections – In-Situ Con TROLL PRO Installation Manual (purchased after 10 May 2012) User Manual

Page 18

Advertising
background image

18

Figure 4.2 AC circuit board and terminal connections

1. Disconnect power to the controller.

2. Use a Phillips screwdriver to remove the controller lid.

3. Use a Phillips screwdriver to remove the AC board cover.

4. Carefully pull the cover out of the enclosure.

5. Connect the green Protective Earth wire to the metal screw as shown in Figure 4.1.

6. Connect the opposite end of the Protective Earth wire to the circuit board as

described in the AC Wiring Descripton Table and Figure 4.2.

7. Connect the stripped-and-tinned black and white conduit or power cord wires to the

terminal strip as shown in the AC Wiring Descripton Table and Figure 4.2.

8. Place the cover back in the enclosure and retighten the screw.

9. Replace the controller lid.

10. Reconnect power to the device.

Relay Connections

Danger—Make sure that power to the controller is disconnected before making

any wiring connections.

Danger—Do not connect low-voltage circuits (<50 V) to the terminal connectors

on the AC board!

Connexions Aux Relais

Danger—Débrancher toute alimentation à l'appareil avant de connecter les fils.

Advertising
This manual is related to the following products: