Sensor wiring connections – In-Situ Con TROLL PRO Installation Manual (purchased after 10 May 2012) User Manual

Page 21

Advertising
background image

21

1. Disconnect power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

3. Connect the stripped-and-tinned PLC wires to the terminal strip as follows.

Terminal Block

Description

GND

Signal ground

485A-

RS485 negative

485B+

RS485 positive

POWER

Input/Output (9-36 VDC)

4. Replace the controller lid and tighten screws.

5. Reconnect power to the device.

Sensor Wiring Connections

Danger—Make sure that power to the instrument is disconnected before making

any wiring connections.

Connexions Pour Le Racord Des Sondes

Danger—Débrancher toute alimentation à l'appareil avant de connecter les fils.

Before attaching the In-Situ sensors, make sure that their
communication settings are set to the factory default values. If
you have a new sensor, these are already set. If you are using an
existing sensor, use Win-Situ 5 Software to reset the
communication settings.

1. Disconnect power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

3. Connect the stripped and tinned wires from the RuggedCable to the terminal strip as

follows.

Terminal Strip

Wire

GND

Black

485A-

Green

485B+

Blue

POWER

Red

Green earth ground
screw

Shielded wire (silver color)

4. Clip or cap the unused wires (brown and white).

5. Replace the controller lid and tighten screws.

6. Reconnect power to the device.

Advertising
This manual is related to the following products: