7 maintenance, Cleaning the outside of the enclosure, Replacement desiccant – In-Situ Con TROLL PRO Installation Manual (purchased after 10 May 2012) User Manual

Page 27: Replacement clock battery, User-serviceable parts, Maintenance

Advertising
background image

27

Maintenance

Danger—Only properly trained and qualified personnel should install the Con

TROLL PRO described in this manual. This instrument should be installed for use in
nonhazardous locations only.

Danger—Installation de l'appareil “controller” décrit dans le guide d'utilisateur doit

impérativement se faire par des personnes qualifiées. L'installation de cet appareil est
seulement prescrite pour des emplacements sans risques.

Cleaning the Outside of the Enclosure

Wipe the enclosure periodically with a damp, soft cloth. Do not use solvents.

Replacement Desiccant

Desiccant ensures the longevity of the electronic components inside the enclosure.
Replace the desiccant (PN 0087630) when it turns from blue to pink.

1. Disconnect AC power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

3. Replace the pink desiccant with new a desiccant pack.

4. Replace the controller lid and tighten screws.

5. Reconnect power to the device.

Replacement Clock Battery

1. Disconnect AC power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

3. Replace depleted clock battery as needed with 3 V MnO

2

-Li battery, CR2032, or

equivalent.

4. Dispose of spent battery properly.

5. Replace the controller lid and tighten screws.

6. Reconnect power to the device.

User-Serviceable Parts

This device contains no user-serviceable electronic parts.

For information regarding service or returns, contact the In-Situ Customer Service
Technicians.

U.S.A. and Canada at 1-800-446-7488, option 3

Internationally at 1-970-498-1500

In-Situ Inc. Attn: Customer Service Department

221 E. Lincoln Ave.

Fort Collins, CO 80524

U.S.A.

Advertising
This manual is related to the following products: