Figure 4-2. message structure, 2 keyword separator, 3 query indicator – KEPCO BIT 232F User Manual

Page 27: Keyword separator -5, Query indicator -5, Message structure -5

Advertising
background image

BIT 232 022800

4-5

FIGURE 4-2. MESSAGE STRUCTURE

You must use the rules above when using keywords. Using an arbitrary short form such as
ENABL for ENAB (ENABLE) or IMME for IMM (IMMEDIATE) will result in an error. Regardless
of which form chosen, you must include all the letters required by that form.

To identify the short form and long form in this manual, keywords are written in upper case let-
ters to represent the short form, followed by lower case letters indicating the long form (e.g.,
IMMediate, EVENt, and OUTPut). The parser, however, is not sensitive to case (e.g., outp,
OutP, OUTPUt, ouTPut, or OUTp are all valid).

4.4.4.2

KEYWORD SEPARATOR

If a command has two or more keywords, adjacent keywords must be separated by a colon (:)
which acts as the keyword separator (e.g., CURR:LEV:TRIG). The colon can also act as a root
specifier (paragraph 4.4.4.7).

4.4.4.3

QUERY INDICATOR

The question mark (?) following a keyword is a query indicator. This changes the command into
a query. If there is more than one keyword in the command, the query indicator follows the last
keyword. (e.g., VOLT? and MEAS:CURR?).

CURR:LEV 3.5;:OUTP ON;:CURR?<NL>

MESSAGE TERMINATOR

KEYWORD

QUERY INDICATOR

ROOT SPECIFIER

MESSAGE UNIT SEPARATOR

DATA

KEYWORD

KEYWORD

KEYWORD SEPARATOR

KEYWORD

MESSAGE UNIT SEPARATOR

DATA

MESSAGE UNIT

DATA SEPARATOR

DATA SEPARATOR

ROOT SPECIFIER

Advertising
This manual is related to the following products: