BOP-1K 031912

B-37/(B-38 Blank)

-311,”Memory Error”

Device Error bit 3

(1)

*SAV (save), *RCL (recall), MEMory:SAVE,
MEMory:UPDAte or CALibrate:STORe error

-314,”Save/recall Memory Lost”

Device Error bit 3

(1)

A memory location that was saved earlier was lost due to a
flash error in Analog processor memory.

-340,”Calibration Failed”

Device Error bit 3

(1)

The calibration results were invalid. This normally indicates
an incorrect calibration caused by operator error such as
performing a current calibration without a load,8 or a voltage
calibration with a load.

-350,”Queue Overflow”

Device Error bit 3

(1)

Error queue was full, error events have been lost.

-363,”Input buffer overrun”

Device Error bit 3

(1)

The input from the controlling computer was longer than 253
characters in length.

-400,”QueryError”

Query Error bit 2

Data requested was lost due to 253 character limit or BIT
4886 output buffer

-420,”Query Unterminated”

Query Error bit 2

Controller sent query but did not accept data from power
supply. See VOLT? (PAR. B.87), CAL:SAVE (PAR. B.87)
commands

TABLE B-5. ERROR MESSAGES (CONTINUED)

ERROR MESSAGE

ESR ERROR BIT SET

(SEE PAR. A.5)

EXPLANATION

(1)