CONTROLLING THE OUTPUT USING EXTERNAL (ANALOG) SIGNALS (REMOTE)

How do I control the mode using an external signal?

$

, highlight External Mode,

!

, highlight Enable,

$

, then

%

to apply

changes and exit. This setting is not saved for power-up (PAR. 3.4.2). Select mode by applying either logic 1 or open circuit for voltage, or
logic 0 or short circuit for current, to pin 2 referenced to pin 9 of the Analog I/O port (PAR. 3.4.2).

How do I use an analog voltage to control the output (main channel)?

$

, highlight Reference input,

!

, highlight External

then

$

, then

%

to apply changes and exit. This setting is not saved for power-up. Apply ±10V signal to pin 11 referenced to pin 4 of the

Analog I/O port to control active channel (voltage or current) (PAR. 3.4.3).

How do I use an analog voltage to control the protection channel?

$

, highlight Protection Limit,

!

, highlight External then

$

, then

%

to apply changes and exit (not saved for power-up). Apply +1V to +10V signal to the following pins referenced to pin 12 of the

Analog I/O port to control limit channel from 10% to full scale: –Current Protect: pin 5, +Current Protect: pin 13, –Voltage Protect: pin 6,
+Voltage Protect: pin 14 (PAR. 3.4.4).

How do I choose between local, digital and analog control of the output?

The unit normally powers up in Local mode. Send-

ing a command via RS 232 or GPIB interfaces automatically puts the unit into Remote mode. Press

!

to restore Local mode. If an analog

voltage is controlling the output (see above), local control of the main channel is disabled and digital commands are not accepted.

How do I use the BOP as an amplifier?

See above to control the main channel and/or protection channel using analog signals. The

nominal (inherent) voltage gain of each model is calculated by (G

NOM-V

) = E

Onom

/ E

REF

. and nominal (inherent) current gain is (G

NOM-I

) =

I

Onom

/ E

REF

where E

REF

= 10V (see PAR. 3.4.3.1).

How do I control the gain of the BOP when used as an amplifier?

$

, highlight Reference input,

!

, highlight ExtlRefLevl then

$

, then

%

to apply changes and exit (not saved for power-up). Apply ±10V signal to pin 11 referenced to pin 4 of the Analog I/O port.

Main channel setting (SET) determines gain (G) by G = SET/10 (PAR. 3.4.3.2).

CONTROLLING THE OUTPUT USING DIGITAL INTERFACE (REMOTE)

How do I use the GPIB interface to control the power supply?

%

,

!

to view current GPBI address. To change GPIB address,

highlight GPIB address,

!

, use keypad to enter new address then

$

. Press

%

to exit or

$

to save for power-up (see PAR. 3.5.3.1).

How do I use the RS 232 interface to control the power supply?

%

,

!

to view. To change Baud Rate, XON/XOFF or Prompt

if needed: highlight Baud Rate, XON/XOFF or Prompt setting,

!

, highlight desired selection, then

$

. When complete,

%

to exit or

$

to save for power-up (see PAR. 3.5.3.2).

How do I change between SCPI and CIIL Data Format?

%

,

!

, to view. To change, highlight Data Format setting,

!

, highlight

SCPI or CIIL then

$

.

%

to exit or

$

to save for power-up. (PAR. 3.5.3.1.2)

How do I change the Device Clear from SCPI to MATE compliant?

%

,

!

, to view. To change, highlight Device Clear setting,

!

, highlight SCPI or MATE then

$

.

%

to exit or

$

to save for power-up. (PAR. 3.5.3.1.2)

FEATURES

How do I save the power supply settings for future use?

!

, highlight Empty location,

!

. The power supply settings (Mode,

type and values for Main Channel Reference and Protection, and output on/off) are displayed. To save a different setting than that dis-
played, highlight the parameter,

!

highlight the selection or enter value using keypad, then

$

.

%

to exit or

$

to save for power-up.

(PAR. 3.3.8)

How do I modify saved power supply settings for future use?

!

, highlight desired location,

!

. The power supply settings

(Mode, Main Channel Reference and value, Protection type and values and output on/off) are displayed. To change a setting, highlight the
parameter,

!

highlight the selection or enter value using keypad, then

$

.

%

to exit or

$

to save for power-up. (PAR. 3.3.8)

How do I recall previously saved power supply settings?

!

, highlight desired location.

@

.to execute stored settings or

!

to

see details, then

@

to program settings. (PAR. 3.3.8)

How do I view or execute a waveform?

@

highlight Waveform name,

!

to view, press STANDBY to turn output on,

$

to exe-

cute.

%

to stop waveform execution. (PAR. 3.3.9)

How do I create a waveform?

Use arrow keys to highlight Empty,

@

,

!

. Enter name using ADJUST or keypad (multiple presses for

letters) and

T

, then

$

. To change Mode highlight selection,

!

.highlight Voltage or Current,

$

to save for power-up. To change pro-

tection, highlight setting,

!

, enter new setting using keypad,

$

. To change count, highlight setting,

!

, enter new setting using keypad,

$

. Define first segment (Type, Frequency, Amplitude (p-p) and Offset) by highlighting parameter,

!

, highlight desired setting or use key-

pad to enter value, then

$

. Highlight segment or End of Segments,

@

to insert next segment prior to highlight or

#

to erase segment.

Define new segment or re-edit existing segment in the same way (including start/stop angle for sine or triangle), then

$

. (PAR. 3.3.9.8)

See above to execute waveform.

How do I calibrate the unit?

%

, highlight Calibration,

!

, then proceed per screen prompts.

How do I configure unipolar operation?

Set either + or – Maximum to zero as follows:

%

, highlight Max/Min Settings,

!

. Higlight

+Voltage Max, -Voltage Min, +Current Max or -Current Min as desired,

!

, press CLEAR to set to zero, then

$

.

%

to exit or

$

to save

for power-up (see PAR. 3.3.4).

Can I further customize the configuration?

Yes, contact Kepco for further information.