Technical data – KROHNE OPTISONIC 7300 EN User Manual

Page 100

Advertising
background image

8

TECHNICAL DATA

100

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Low-flow cutoff

Low-flow cutoff

Low-flow cutoff

Low-flow cutoff

On

0...±9.999 m/s; 0...20.0%, settable in 0.1% steps, separately for each current

and pulse output.

Off

0...±9.999 m/s; 0...19.0%, settable in 0.1% steps, separately for each current

and pulse output.

Time constant

Time constant

Time constant

Time constant

Function

Can be set together for all flow indicators and outputs, or separately for:

current, pulse and frequency output, limit switches and the 3 internal

counters.

Time setting

0…100 seconds, settable in 0.1 second steps.

Current input

Current input

Current input

Current input

Function

For conversion to standard conditions, input from external temperature and

pressure transmitters is required.

Operating data

Basic I/Os

Modular I/Os

Ex i

Active

-

U

int

= 24 VDC

I ≤ 22 mA

I

max

≤ 26 mA

(electronically limited)

U

0, min

= 19 V at

I ≤ 22 mA

U

int

= 20 VDC

I ≤ 22 mA

U

0, min

= 14 V at

I ≤ 22 mA

No HART

®

No HART

®

U

0

= 24.1 V

I

0

= 99 mA

P

0

= 0.6 W

C

0

= 75 nF / L

0

= 0.5 mH

No HART

®

Passive

-

U

ext

≤ 32 VDC

I ≤ 22 mA

I

max

≤ 26 mA

(electronically limited)

U

0, max

= 5 V at

I ≤ 22 mA

U

ext

≤ 32 VDC

I ≤ 22 mA

U

0, max

= 4 V at I ≤ 22 mA

No HART

®

No HART

®

U

I

= 30 V

I

I

= 100 mA

P

I

= 1 W

C

I

= 10 nF

L

I

=0 mH

No HART

®