8 command #138 read analog output function – KROHNE UFC 500 HART User Manual

Page 22

Advertising
background image

HARTô Smart Communications Protocol

UFC500 Command Specification Rev: 2.0 Release Date: 9 Jun 2001

© 2001 KROHNE Mefltechnik D-47058 Duisburg

Page

22

8.8

Command #138

Read Analog Output Function

Reads the function of current/pulse output.

REQUEST DATA BYTES

#0
ANALOG
OUTPUT
NUMBER
CODE

RESPONSE DATA BYTES

#0

#1

ANALOG

ANALOG

OUTPUT

OUTPUT

NUMBER

FUNCTION

CODE

CODE

Data Byte #0

Analog Output Number Code, equals м1о for current output and м2о - for pulse output

Data Byte #1

Analog Output Function Code, depends on the output type:

For current output:
0 - Off.
1 - Direction indication.
2 - 1 Dir, output traces flowrate measurements only in one (positive) direction.
3 - Forward/reverse flow, the complete flowrate dynamic range (PV Upper Range Value +
Reverse Range) is mapped onto the range of the output current. Refer also to the instrument
Operating Manual and to Commands #140, Read Current Output Parameters, #158, Read
Transmitter Variable Range Values.
4 - 2 Dir, output traces flowrate measurements in both directions, for every direction the
corresponding flowrate dynamic range (PV Upper Range Value for positive direction and
Reverse Range - for negative) is mapped onto the range of the output current. Refer also to
the instrument Operating Manual and to Commands #140, Read Current Output Parameters,
#158, Read Transmitter Variable Range Values.
5 - Transit time.

For frequency/pulse output:
0 - Off.
1 - Error indication.
2 - 1 Dir, output traces flowrate measurements only in one (positive) direction.
3 - 2 Dir, output traces flowrate measurements in both directions.
4 - Transit time.

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1

Undefined

2

Invalid Selection (of Analog Output)

3-4

Undefined

5

Too Few Data Bytes Received

6

Local Device User

7-127

Undefined

Advertising