KROHNE UFC 500 HART User Manual

Page 41

Advertising
background image

HARTô Smart Communications Protocol

UFC500 Command Specification Rev: 2.0 Release Date: 9 Jun 2001

© 2001 KROHNE Mefltechnik D-47058 Duisburg

Page

41

8.22

Command #156

Read Device-Specific Units And Enumerators

Reads device-specific units and controls.

REQUEST DATA BYTES

NONE

RESPONSE DATA BYTES

#0

#1

#2

#3

PLAUSIB

NUMBER

NUMBER

WEIGHT

ERROR

PLAUSIB

PLAUSIB

POINT

ERRORS

ERRORS

OK

MSB

LSB

#4

#5

PULSE

PULSE

VOLUME

RATE

UNITS

UNITS

#6

#7

#8

#9

REV

DISPLAY

ERROR

DISPLAY

SCALE

TOTAL

MESSAGE

TRANSIT

CONTROL

CONTROL

CONTROL

TIME
CONTROL

Data Byte #0

Error limit in % of measured value for plausibility statement.

Data Byte #1..#2

Limit value for the counter of incorrect measurements.

Data Byte #3

Weight factor for correct measurements.

Data Byte #4

Pulses per volume units, used only for scaling of PulseVolume parameter (refer to Command
#142, Read Frequency/Pulse Output Parameters) and do not affect physical output. For unitsí
codes refer to Table 9.3.

Data Byte #5

Pulse rate units, used only for scaling of PulseRate parameter (refer to Command #142, Read
Frequency/Pulse Output Parameters) and do not affect physical output. For unitsí codes refer
to Table 9.4.

Data Byte #6

Reverse scale control, defines whether the Upper Range Value for reverse flow is the same as
for forward, or different:
0 - Different,
1 - Same as for forward flow.
In the first case the Upper Range Value for reverse flow can be accessed via Command
#158, Read Transmitter Variable Range Values.

Data Byte #7

Display totalizer control, affects the local display of the instrument and the one of the HART
communicator HC275:
0 - "+Totalizer",
1 - "-Totalizer",
2 - "+/-Totalizers",
3 - "Totalizers' sum",
4 - "All totalizers",
5 - "No display".