Earth Voltage Measurement

Set the range switch to EARTH VOLTAGE position in the condition of

q

. Earth voltage will be indicated on the display. Make sure that the

voltage is 10V or less.

When the display reads more than 10V, it may result in excessive

errors in earth resistance measurement. To avoid this, make

measurement after reducing the voltage by turning off the power

supply of the equipment under test etc.

e

Simplified Measurement

Set the range switch to 2000

position and press the test button. LED

remains illuminated during testing. Turn the range switch to 200

and

20

when the earth resistance is low. This indicated value is the earth

resistance of the earthed equipment under test.

Note :

If the auxiliary earth resistance of auxiliary earth spike C is too

high to make measurement, the display reads ' . . . '. Recheck the

connection of each test lead and the earth resistance of auxiliary

earth spike.

r

Simplified Measurement Value

Two-terminal method is used for simplified measurement. In this

method, earth resistance value re of earth electrode connected to

terminal P is added to true earth resistance value Rx and shown as

an indicated value Re.

Re = Rx + re

If the re is known beforehand, true earth resistance value Rx is

calculated as follows.

Rx = Re --- re

---- 13 ----

Note :

When the simplified measurement probes are not used, short P

and C terminals.

DANGER

Please be sure to use a voltage detector to check a common earth of

commercial power supply.

Do not use the instrument to check a common earth of commercial

power supply.

A danger will be caused because the voltage may not be displayed

even in case of a live conductor when the connection of the earth

electrode to be measured has come off, or when the connection of the

test leads of the instrument is not correct etc.

Do not use the instrument to measure the voltage of commercial power

supply.

The instrument is not designed for voltage measurement of commercial

power supply. When using the adjunctive simplified measurement

probe MODEL 7127, P and C terminals will be short-circuited and the

input impedance will be reduced. The residual current circuit breaker

may operate when making measurement of the voltage in the circuit

with the breaker.