KYORITSU 4300 User Manual

Page 4

Advertising
background image

̶ 2 ̶

#

DANGER

● Never make measurement on a circuit in which the electrical

potential exceeds 300V.

● Do not attempt to make measurement in the presence of

flammable gasses. Otherwise, the use of the instrument may

cause sparking, which can lead to an explosion.

● Never attempt to use the instrument if its surface or your hand

is wet.

● Be careful not to short-circuit the power line with the metal

part of the test leads when measuring voltage. It may cause

personal injury.

● Do not exceed the maximum allowable input of any

measurement range.

● Never open the Battery compartment cover during a measurement.

#

WARNING

● Never attempt to make any measurement if any abnormal

conditions, such as a broken cover or exposed metal parts are

present on the instrument and test leads.

● Do not press the Test button when connecting the test leads to

the instrument.

● Do not install substitute parts or make any modification to

the instrument. Return the instrument to your local Kyoritsu

distributor for repair or re-calibration in case of suspected faulty

operation.

● Do not replace batteries if the instrument is wet.

● Ensure that the L-shaped Banana plug of MODEL7248 is firmly

inserted into the E terminal.

● Power off the instrument when opening the Battery

compartment cover for battery replacement.

#

CAUTION

● Ensure that the instrument is powered off after use and test

leads are disconnected. Remove the batteries if the instrument

is to be stored and will not be in use for a long period.

● Do not expose the instrument to direct sunlight, high

temperature, humidity or dew.

● Use a damp cloth with neutral detergent or water for cleaning

the instrument. Do not use abrasives or solvents.

● Do not store the instrument if it is wet.