KYORITSU 6010B User Manual

Page 10

Advertising
background image

― 8 ―

● Operating Error of RCD (IEC 61557-6)

The influencing variations used for calculating the operating error are denoted as follows:

Temperature : 0℃ and 35℃
Earth electrode Resistance (shall not exceed below) :

System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V

Instrument dimensions:-

175 X 115 X 86mm

Instrument weight:-

840g including batteries.

Reference conditions

Specifications are based on the following conditions
except where otherwise stated:-

1.Ambient temperature: 23± 5:
2.Relative humidity 45% to 75%
3.Position: horizontal
4.AC power source 230V, 50Hz
5.DC power source: 12.0 V, ripple content 1% or

less

6. Altitude up to 2000m, Indoor use

Battery type

Eight R6 or LR6 batteries.

Low battery warning

"

" symbol appears in the display if the battery

voltage drops below 8V.

Operating temperature

0 to +40:, relative humidity 80% or less, no

and humidity.

condensation.

Storage temperature

-20 to +60℃, relative humidity 75% or less, no

and humidity

condensation.

Surge protection

transient overvoltage 4000V

Insulation resistance

greater than 50MΩ at 1000V DC
(between enclosures and electrical circuits)

Function

-10% - 0%

0% -+10%

-10% -+10%

Operating error of trip current

Ч1/2
Ч1, FAST
Auto Ramp

Earth electrode resistance (

Ωmax.)

I

Δn (mA)

UL50V

2000

600
200
130

80

10
30

100
300
500

2000

600
200

65
40

UL25V

Advertising