KYORITSU 6010B User Manual

Page 11

Advertising
background image

LED indication of live

Illuminates if there is an alternating voltage of 20V

circuit warning

AC or more in the circuit under test before Continuity
or Insulation resistance tests. When DC voltage is
detected across the measuring terminal the LED
lights up.

LED indication of

The P-E and P-N LEDs illuminate when the wiring of

correct polarity

the circuit under test is correct. The reverse LED
"

" is lit when P and N are reversed.

Display

The liquid crystal display has 3 1/2 digits with a
decimal point and units of measurement (Ω, MΩ, V,
mA and ms) relative to selected function.

Overload protection

The continuity test circuit is protected by a 0.5 A 600
V fast acting (HRC) ceramic fuse mounted in the
battery compartment, where a spare fuse is also
stored.
The insulation resistance test circuit is protected by
a resistor against 1200 V AC for 10 seconds.

Mains Voltage

On connecting test leads to the circuit under test at

Indication

Loop, RCD and Uc ranges, the LCD reads VL-PE.
Indications are as follow:
Less than 100V : "Lo v"
100V~259V

: voltage value and "

"

260V~300V

: voltage value and "Hi v"

alternately, and "

"

Over than 300V : "Hi v" and "

"

― 9 ―