― 32 ―

11. GENERAL

The test button can be locked down for ease of use by pressing it and turning
clockwise. Do not forget to release test button by turning it counterclockwise
before disconnecting the instrument from the test points. Failure to do so may
leave the tested circuit in a charged condition when carrying out insulation test.
The instrument is provided with a sliding cover to ensure that leads for testing
continuity and insulation resistance cannot be connected at the same time as
test leads for Loop/RCD/Uc testing. If this sliding cover is damaged so that it
fails to perform its function, do not use the instrument and return it to your
distributor for attention.

12. SERVICING

If this tester should fail to operate correctly, return it to your distributor stating
the exact nature of the fault. Before returning the instrument ensure that:-

1.The leads have been checked for continuity and signs of damage.
2.The continuity mode fuse (situated in the battery compartment) has been

checked.

3.The batteries are in good condition.

Please remember to give all the information possible concerning the nature of
the fault, as this will mean that the instrument will be serviced and returned to
you more quickly.