SHIMPO TNP-Series User Manual

Page 13

Advertising
background image

12







Resetting the Peak values to Zero


MEAN is real time value based on the average data captured in 1000 samples
per second.


The maximum display update for all modes is 8 times/second.
This update rate can be adjusted by changing F02 from function mode
(See section VII - Values available 1, 2, 4, 8 times/second).

Comparator Function

This compares the measured value to the upper and lower limits entered under
function mode for HI and LO limits.
If both HI and LO limits are set to “0000” this feature is not available
(deactivated).
The following conditions are valid under comparator mode:

• HI>LO
• HI=LO (HI Red LED indicator will be lit on the Comparator LED indicators).

Utilizing this feature makes it an ideal tool for quality assurance checks.

Example: HI is set to 100 and LO is set to 50. Based from the conditions met
the corresponding LED will light up (HI (Red LED), GO (Green LED), LO (Red
LED)).










From the previous example any values greater than 100 will light up the HI red
LED.
Any values lower than 50 will light up the LO red LED.
Values measured in between these values (100<X<50) will give a GO green LED
indicator.

100

Upper limit value

50

Lower limit value

HI:ON

LO:ON

0

OK:ON

Test 1 = 105 HI
Red LED
Test 2 = 45 LO
Red LED
Test 3 =65, GO
Green LED

Open

Mean

Close

Open

Mean

Close

Nm

Nm

Zero/RST

Advertising