SHIMPO TNP-Series User Manual

Page 8

Advertising
background image

7



If the display shows the normal operation of the torque meter, and no
F01 is seen on the display, the Zero/RST key was released earlier than
what is required. Turn off the torque meter and repeat the process.

VII. Function Mode Table

FUNCTION

Sub display

Options

Initial setting

Measuring unit

F01

Changes units of measure
TNP-0.5 :

mN-m, N-cm, g-cm, lb-in

TNP-2, 5, 10 :

N-m, N-cm, Kg-cm, lb-in

TNP-0.5 : mN-m

TNP-2, 5, 10 : N-m

Display-(update rate)

F02

Switch 1, 2, 4, 8 times/second

2

Auto power OFF

F03

10 minutes or on

10 minutes

Sign Orientation

F04

-0000 CCW (Open "-");
0000 CCW(Open "+")

0000

Upper comparator

value (Hi limit)

HI

0000 – 9999: with decimal
point. (Setting the values to
zero disables this function) .

0000

Lower comparator

value (Lo limit)

LO

0000 – 9999: with decimal
point (Setting the values to
zero disables this function).

0000

How to change value on the function selected?

To change the values of the function selected, use the Mode key button to scroll
through the options and the Zero/RST button to move to the next Mode.


Pressing the Mode button after setting the HI/LO limits exits out
of the function mode. You will need to reenter the settings to
change additional values.