Waterproof ECScan Low Instructions

68X247754 11/00 Rev 0

Conductivity Testing

1. Remove electrode cap. Switch unit on (ON/OFF

key).

2. Dip electrode into test solution. Make sure

sensor is fully covered.

3. Wait for reading to stabilize (Automatic

Temperature Compensation cor rects for
temperature changes). Note reading.

4. Press ON/OFF to turn off tester. Replace

electrode cap. Note: Tester automatically
shuts off after 8.5 minutes of non-use.

HOLD Function

Press HOLD key to freeze display. Press HOLD again
to release.

Tester Maintenance

• To improve performance, clean the electrode

by rinsing them in alcohol for 10 to 15 minutes.

• Replace all batteries if low battery indicator

appears, or if readings are faint or unstable.

• If you experience drift, periodically let electrode

fully dry.

When you need a new electrode, see “Electrode
Replacement” on insert in back of box.

Self-Diagnostic Messages

bAt Weak batteries–replace with fresh ones
specified by manufacturer.

Err Wrong keypress.

Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.

---- Calibration mode activated but not performed.

Before You Begin

Remove electrode cap. Soak electrode for a few

minutes in alcohol to remove oils.

Calibration

Tester is factory calibrated. However, to ensure
accuracy, calibrate conductivity on a regular basis.

Select a calibration standard appropriate for the

ECScan Low tester (between 200 µS and

1990 µS). It is best to select a standard close

to the test solution value.

1. Open battery compartment lid (end with

lanyard loop).

2. Orientate the battery compartment as

shown in Diagram 1.

3. Rinse electrode in deionized water, then

rinse it in calibration standard, then dip it

into a container of calibration standard.

4. Switch unit on (ON/OFF key). Wait

several minutes for display to stabilize.

5. Press the INC or DEC key to adjust

reading to match the calibration

standard value.

6. After 3 seconds without a key

press, the display flashes 3 times,
then shows ‘ENT’. The tester accepts
calibration value; returns to
measurement mode.

7. Replace battery cap.