Electrode replacement, Applications – Thermo Fisher Scientific ECScan High/Low & TDScan High/Low User Manual

Page 9

Advertising
background image

A

B

REMOVE THE OLD ELECTRODE

REPLACE THE ELECTRODE

(Picture B)

(Picture A)

Electrode Replacement

You can replace the electrode module at
the fraction of the cost of a new tester.
When the tester fails to calibrate or gives
fluctuating readings in calibration
standards, you need to change the
electrode.

1.

With dry hands, grip the ribbed tester
collar with electrode facing you. Twist the
collar counter clockwise (see picture A).
Save the ribbed tester collar and O-ring
for later use.

2.

Pull the old electrode module away from
the tester.

3.

Align the four tabs on the new module
so that they match the four slots on the
tester (see picture B).

4.

Gently push the module onto the slots
to sit it in position. Push the smaller O-
ring fully onto the new electrode module.
Push the collar over the module and thread
it into place by firmly twisting clockwise.

Water quality testing • water and

Wastewater treatment • hydroponics

• labs • boilers • factories

• ecology studies • and more!

Applications

Tester

TDScan TDScan

ECScan ECScan

Low

High

Low

High

Range

0 to

0 to

0 to

0 to

1990

10.00

1990

19.90

ppm

ppt

µS/cm

mS/cm

Resolution

10

0.10

10

0.10

ppm

ppt

µS/cm

mS/cm

Accuracy

±1% Full Scale

TDS factor 0.4 to 1.0 selectable

-

Calibration

200 to

2 .00 to

200 to 2.00 to

Standard

1990

10.00

1990

19.90

Range

ppm

ppt

µS/cm

mS/cm

Calibration

1 point (calibration range is ±30%

of factory default parameter)

ATC

0 to 50 °C

Temp

2% per °C

Coefficient

Operating

0 to 50 °C

Temp

Power

Four X 1.5V

Battery

(Type: A76, LR 44 or V136A);

> 140 hrs continuous use

Dimension;

Tester: 16.5 cm x 3.8 cm; 90 g

Weight

Boxed: 22 cm x6 cm x 5 cm; 170 g

Push-button calibration

Fast, stable, repeatable readings

Available through authorized distributors

S P E C I F I C A T I O N S

Lanyard
connection

Replaceable

Sensor

Twist-off

Battery

Compartment

Manufactured By:

Eutech Instruments Pte Ltd

Blk 55, Ayer Rajah Crescent,

#04-14/24, Singapore 139949

Tel: (65)778 6876 Fax: (65)773 0836

E-Mail: [email protected]

Web-Site: http://www.eutechinst.com

Made in Singapore