TDS Testing

1. Remove electrode cap. Switch unit on (ON/OFF

key).

2. Dip electrode into test solution. Make sure sensor

is fully covered.

3. Wait for reading to stabilize (Automatic

Temperature Compensation cor rects for
temperature changes). Note reading.

4. Press ON/OFF to turn off tester. Replace electrode

cap. Note: Tester automatically shuts off after
8.5 minutes of non-use.

HOLD Function

Press HOLD key to freeze display. Press HOLD again
to release.

Setting TDS Factor

This tester lets you select a TDS factor of 0.4 to 1.0.

1. Open battery compar tment. With meter on,

press the HOLD key, the press the INC key
See Diagram 1.

2. Press the INC or DEC key to adjust the TDS factor.

3. After 3 seconds without a key press, the display

flashes 3 times, then shows ‘ENT’. Tester accepts
TDS factor and returns to measurement mode.

4. Replace battery cap.

Self-Diagnostic Messages

bAt Weak batteries–replace with fresh ones specified
by manufacturer.

Err Wrong keypress.

Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.

---- Calibration mode activated but not performed.

Before You Begin

Remove electrode cap. Soak electrode for a few

minutes in alcohol to remove oils.

Calibration

Tester is factory calibrated. However, to ensure

accuracy, calibrate TDS on a regular basis.

Select a calibration standard appropriate for the

TDScan Low tester (between 200 ppm and

1990 ppm). It is best to select a standard

close to the test solution value.

1. Open battery compartment lid (end with

lanyard loop).

2. Orientate the battery compartment as

shown in Diagram 1.

3. Rinse electrode in deionized water, then

rinse it in calibration standard, then dip it

into a container of calibration standard.

4. Switch unit on (ON/OFF key). Wait

several minutes for display to stabilize.

5. Press the INC or DEC keys to adjust

reading to match the calibration

standard value.

6. After 3 seconds without a key

press, the display flashes 3 times,

then shows ‘ENT’. The tester accepts
calibration value; returns to
measurement mode.

7. Replace battery cap.