Mains (line) supply, 20 - 50v ac/dc supply, Process input – West Control Solutions N8080 (for 2xxx) User Manual

Page 22: Digital input option

Advertising
background image

1.3.1 Mains (Line) Supply

The Indicator will operate on 90 - 264V AC 50/60Hz mains (line) supply. The power

consumption is approximately 4 watts.

CAUTION: This equipment is designed for installation in an enclosure

which provides adequate protection against electric shock. Local

regulations regarding electrical installation should be rigidly

observed. Consideration should be given to prevention of access to

the power terminations by unauthorised personnel. Power should be

connected via a two-pole isolating switch (preferably situated near

to the equipment) and a 1A fuse, as shown in Figure 1-4.

If the Indicator has relay outputs in which the contacts are to carry mains (line)

voltage, it is recommended that the relay contact mains (line) supply should be

switched and fused in a similar manner but should be separate from the Indicator

mains (line) supply.

1.3.2 20 - 50V AC/DC Supply

Power should be connected via a two-pole isolating switch and a 315mA

slow-blow (anti-surge Type T) fuse. With this option fitted, the Indicator will accept

20 - 50V AC @ 50/60Hz or 20 -50V DC in the polarity shown in Figure 1-4.

1.3.3 Process Input

The process input will accept millivolt, volt and milliamp DC ranges.

1.3.4 Digital Input Option

Terminals 16 and 17, when this option is fitted, may be used for either of two

functions (selectable in Configuration Mode); (a) Tare facility or (b) Security facility.

These terminals may be connected to (a) the voltage-free contacts of a switch or

relay, or (b) a TTL-compatible voltage. With the required function selected in

Configuration Mode (see Subsection 3.4), operation is as follows:

1-4

Volume II

PM091-1

59136

Voltage Free Operation

TTL-conpatible

Operation

Tare Facility

Security Facility

Contacts open

Signal >2.0V

Current process

variable value used

as new “zero” point

to create an

automatic offset.

Entry into Program

Mode prohibited

Contacts Closed

Signal <0.6V

No automatic offset

applied.

Entry into Program

Mode permitted