The Tank Level tab

The

Tank Level

tab allows you to setup the gauge as a height or percent of capacity indicator for a tank or reservoir by entering in a few basic measurements.

The tank level assumes the fl uid density is constant, and that the fl uid is not compressible. You will need the following information:

Desired units to be displayed on gauge (feet, meters, % of capacity, etc.). If % of capacity is chosen, you will also need the height of the tank.

The vertical distance from the gauge to the bottom of the tank, and if the gauge is above or below the bottom of the tank.

Media density, in kg/m³, lbs/ft³, or slugs/ft³.

The local gravity. If local gravity is unknown, you can calculate local gravity if the latitude and elevation is known, or use the standard gravity value of
9.80665.

The User Factor and User Off set will be calculated from these parameters.

Note:

The User Off set and User Unit Wizard is only available for gauges running fi rmware version R0013 or later (for XP2i gauges) or R0113 or later

(for XP2i-DD gauges). Use

XP2iUpdate

to upgrade any gauge to the latest fi rmware.