CALIBRATION OPTIONS

Userspan:

Usually set to 1 at the factory, and used to adjust the gauge for component aging. All readings are multiplied by this number. For instance, if the

readings are low by 0.1%, set this number to 1.001 to correct the readings (1.001 is equivalent to 100.1%).

Disable Keypad Userspan:

Checking this box prevents calibration adjustments using the keypad, only. This does not aff ect the ability to change the users-

pan value with Confi gXP.

Zero Value Adjust (available on Absolute gauges only):

Allows long term zero drift of absolute gauges to be calibrated out. Zero value adjust is added to

the displayed pressure.

AS RECEIVED/AS LEFT REPORT

The As Received / As Left Report is a summary of the gauge confi guration before and after any changes made using Confi gXP. The As Received state is the

condition of the gauge after the initial connect. The As Left state includes all changes since the initial connect, including any changes not yet written to the

gauge.