Crystal ConfigXP Digital Pressure Gauge Software User Manual

Page 9

Advertising
background image

Settings

9

Confi gXP User’s Manual

Using the User Units Wizard

The User Units Wizard operates in two modes: Linear Fit or Tank Level. Both modes calculate an optimum User Factor and User Off set.

The Linear Fit tab

The

Linear Fit

tab allows you to quickly calculate the best User Factor and User Off set by entering actual measurements and the corresponding desired read-

ing for each measurement. In each row, enter the displayed pressure in the fi rst column and the reading you want the gauge to display in the

Desired Indica-

tion

column. The User Factor and User Off set will be recalculated after each point entered. The

Gauge Reading

column shows what the gauge will actually

display using the calculated User Factor and User Off set. The

Delta

column indicates how much error there is between the Desired Indication and the Gauge

Reading.

Advertising