Module Queries and Commands

4

nVision Programming Instructions

Module Queries and Commands

4

X

Serial number

SN?

. . . . . . . . . . The serial number is returned as a six digit number.

commands

X

automatic Shutoff timer

AO!XXXX

Set the auto off time for the chassis. (3600 seconds is the max auto off time.)

module QuerIeS and commandS

Communication with the module requires you to specifically identify the module location. As shown below, this is done by appending a 1, 2, or 3 to the end
of the query.

Queries

X

Firmware Version

MOD:VER? 1 ...

Returns the nVision firmware version of the lower module in the RXXXXXX.XX format.

MOD:VER? 2 ...

Returns the nVision firmware version of the upper module in the RXXXXXX.XX format.

MOD:VER? 3 ...

Returns the nVision firmware version of the BARO module in the RXXXXXX.XX format.

X

Full Scale range

MOD:FR?

Note:

This query is not applicable to the RTD100 module. Full scale range for MA20 switch test is “1". Full scale range for BARO is “15".

XXXX

= 0 disables Automatic Shutoff (always on)

XXXX

= seconds:

30

45

60

120

180

240

seconds

seconds

(1 min)

(2 min)

(3 min)

(4 min)

300

600

900

1200

1800

2700

3600

(5 min)

(10 min)

(15 min)

(20 min)

(30 min)

(45 min)

(1 hr)

These values can be set to anything from 1 to 3600 seconds. However, once connected to CrystalControl

this value will be reset to the closest value shown in the table. For example, sending the command AO!75

instructs the nVision to set the Automatic Shutoff time to 1 minute.