background image

Module Queries and Commands

5

nVision Programming Instructions

Module Queries and Commands

5

X

message Store

MOD:MSG? #

X

model

MOD:MOD? #

X

reading

MOD:RD? #

Note:

The MA20 switch test states are “1” for open, and “0” for closed. Units for MA20 module are either “mA, % 4-20mA, %10-50mA

(for MA-20+ modules only), VDC, or Switch Test.”

X

Serial number

MOD:SN? #

X

unit

MOD:UNIT? #

Response:

PSI

,

kg/cm2

,

inHg

,

inH2O

,

mmHg

,

mmH2O

,

kPa

,

bar

,

mbar

,

MPa

,

user

,

C

,

F

,

R

,

K

,

Ohm

,

mA

,

%4-20mA

,

%10-50mA

,

VDC

, or

SwitchTest

Note:

This query is only available when the nVision has R080007 or greater firmware.

X

Water reference

MOD:H2O? #

Response:

4C

,

60F

, or

68F

Note:

This query is only available when the nVision has R080007 or greater firmware and is populated with a pressure module (PM).