Specifications – Crystal XP2i 2nd Generation User Manual

Page 14

Advertising
background image

Specifications

12

XP2i Operation Manual

Specifications

accuracy

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

Gauges must be exercised whenever exposed to significant changes in environmental conditions to achieve these specifications, and (if not an absolute
model), rezeroed. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest. A properly exercised gauge
will return to a perfect zero reading (or return to the same ambient barometric reading). Absolute versions of

XP2is have a “B” in front of “XP2I” in the part

number.

Exposure to environmental extremes of temperature, shock and/or vibration may warrant a more frequent recertification period.

20 to 100% of Full scale . . . . . . . . . ±(0.1% of Reading).

0 to 20% of Full scale . . . . . . . . . . . ±(0.02% of Full Scale).

Vacuum, for 300 PsI, 20 bar, 2000 kPa, and lower pressure gauges:

±(0.25% of Full Scale), where F.S. = -14.5 PSIG, -1.0 bar, -99.9 kPa.

Not recommended for continuous use at high vacuum. Refer to

XP2i-dP

data sheet for gauges that are intended for continuous high vacuum use.

temperature

operating . . . . . . . . . . . . . . . . . . . . . . . -10° to 50°C (14° to 122°F).

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

storage . . . . . . . . . . . . . . . . . . . . . . . . . -40° to 75°C (-40° to 67°F).

Battery should be removed if stored for more than one month.