safety Instructions for Hazardous Locations

engLIsH

(English)

Do not use the RS-232 connector in a hazardous location.

Replace batteries in non-hazardous locations, with approved batteries, only.

It is the users responsibility to understand the proper application of this product in potentially explosive atmospheres.

approved Batteries

engLIsH

(English)

The XP2i is Intrinsically Safe only if powered by one of the following battery types:

Rayovac

® Maximum Plus™

Energizer

® E91

Energizer

® EN91

Duracell® MN1500

Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety - do not assume other models are equivalent.

Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.

Instrucciones de seguridad Para Zonas Peligrosas

esPaÑoL (Spanish)

No use el conector RS-232 en zona clasificada.

Cambie las pilas en zona no clasificada, solo con pilas aprobadas.

Es responsabilidad del usario comprender la aplicación de este producto en atmósferas potencialmente explosivas

Pilas aprobadas

esPaÑoL (Spanish)

El XP2i solo es intrínsecamente seguro si se alimenta con uno de los siguientes tipos de pilas:

Rayovac

® Maximum Plus™

Energizer

® E91

Energizer

® EN91

Duracell® MN1500

Se han probado muchos otros tipos de baterías pero han fallado el cumplimiento de los requisitos para la seguridad intrínseca - No asuma que otros modelos
son equivalentes.

Energizer está fabricado por Energizer Holdings, Inc., y por Eveready Battery Company, Inc.