Gain – FUTEK CSG110 (JM-2A / JM-2AD)-Series User Manual

Page 7

Advertising
background image


- 7 -





Tel: (949) 465-0900

Fax: (949) 465-0905

Toll Free: (800) 23 FUTEK

Ex. If you are using your CSG110 with a tension and compression load cell and you
have tension setup as the positive direction, but now you would like to have
compression as the positive direction, all you have to do is flip the DIP switch from the
default polarity position to the reverse polarity position.

UP

reverse

DOWN straight

SW2: Polarity

Gain

There are eight gain settings available on the CSG110: from 0.5 mV/V to 10 mV/V. The
default setting is configured to 2 mV/V. Select the value closest to your input range by
flipping the DIP switches to the appropriate configuration.

Ex. If you are using a 2 mV/V sensor with the 5 VDC excitation, then you would want to
use configuration 2 for 1 mV/V. If you are using a 2mV/V sensor with 10 VDC
excitation then you would want to use configuration 4 for 2 mV/V.