FUTEK IPM500 (D500) Digital Display User Manual

Page 44

Advertising
background image

44

19. TEDS INTERFACE


U

pon power up, a TEDS enabled IPM490 or IPM500 meter will display a sequence of "3333",

"2222", and "1111" followed by the firmware revision, alerting the user of TEDS capability.
Ensure that the proper connections are made from pins 5 and 6 of the TEDS connector to the
DB-9 converter PCB before plugging in a TEDS enabled sensor, as described in Section 4.
Once the sensor is connected, the meter will display the manufacturer's ID and TEDS template
number, followed by an automatic reset. If the sensor contains meter analog output or decimal
point settings, the meter will automatically configure these settings.

TEDS MENU SELECTIONS

The following menu selections must be entered before retrieving TEDS information from your
TEDS enabled sensor. The meter must be configured for 2-point calibration before the sensor
is connected; otherwise analog output settings will not be set up properly.

Press Menu
Select Key

Press Digit
Select Key

Press Value Select
Key

SEtuP

Press

key until

SEtuP is displayed.

00000

Press to select the
4

th

digit to flash.

00010

Press to set digit 4 to 1 for coordinates of
2 points scaling method.

SEr 1

.

Press

key until

SEr 1 is displayed.

000U

Press to select digit
to flash.

150

Set digit 3 to

1

for filtered signal.

150

Set digit 4 to

5

for 9600 baud.

150

Set digit 5 to

0

for 60/s output rate.

SEr 2

.

Press

key until

SEr 2 is displayed.

0000U

Press to select digit
to flash.

0000

Set digit 2 to

0

for no line feed.

0000

Set digit 3 to

0

for no alarm data.

0000

Set digit 4 to

0

for continuous

output.

0001

Set digit 5 to

1

thru

F

for addresses

1 thru 15 or to

0.

thru

F.

(with deci-

mal point) for addresses 16 thru 31.

SEr 3

.

Press

key until

SEr 3 is displayed.

00000U

Press to select digit
to flash.

00000

Set digit 1 to

0

for full duplex.

00000

Set digit 2 to

0

for standard

continuous output.

00000

Set digit 3 to

0

for normal RTS.

00000

Set digit 4 to

0

for termination cha-

racters only at end of all items.

00001

Set digit 5 to

0

for data readings in

continuous mode.