FUTEK IPM500 (D500) Digital Display User Manual

Page 15

Advertising
background image

Baud

Rate

Time

1 Item

Min

Gate

Time

2 Items

Min

Gate

Time

3 Items

Min

Gate

Time

4 Items

Min

Gate

300
600

1200

2400

4800
9600

19200

.37s
.18s

.09s

.05s

.02s
.01s

.01s

.34s
.15s

.06s

.02s

.01s
.01s

.01s

.70s
.35s

.18s

.09s

.04s
.02s

.01s

.67s
.32s

.15s

.06s

.01s
.01s

.01s

1.03s

.52s

.26s

.13s

.07s
.03s

.02

1.00s

.49s

.23s

.10s

.04s
.01s

.01s

1.37s

.68s

.34s

.17s

.09s
.04s

.01s

1.34s

.65s

.31s

.14s

.06s
.01s

.01s

The data transmission rate may be reduced by sending data every other reading, every fourth

reading, or less. This selection is made in “Ser 1”. A computer, if busy with other tasks, may be

unable to keep up with the faster data rates of the meter, so a handshake function is available
that provides the computer with control over the meters’ data transmissions.

7.3 RTS CONTROL

RTS control does not apply to transmitter, where the RTS line is always held high, nor to
RS485. DPMs and counter / timers have two RS232 RTS modes: unlatched and latched.
In the unlatched mode, the measurement transmission is enabled by a high RTS level and is

disabled by a low RTS level. When disabled, any character being sent is completed. When

enabled, any characters remaining in the data format are transmitted before the next measu-

rement transmission. The computer, when its receive buffer is nearly full, takes the RTS line
low to halt data transmission. When its receive buffer has emptied, it takes the RTS line high to

enable more data transmissions. Some measurements could be missed in the process. The

latched and unlatched modes are selected in "config" "digit 2" in the DPM and by “Ser 3” in the
Counter and Scale Meter.
In the latched mode, the RTS input is polled every 3.3 ms. When a high level is detected, RTS

is latched true, even though the RTS line goes low immediately. At the end of each calculation,

the latched RTS value is checked. If it is true, a complete measurement transmission (from 1 to

4 values) is made without interruption, regardless of the state of the RTS line during that time.

At the end of the complete transmission, the latched RTS value is reset false, even though the

RTS line may be high at that instant. The RTS latch does not go true again until the RTS line is
first returned to a low level after the completion of the transmission and then is taken high

again. Latched control provides “print command” operation by sending a transmission for each

RTS pulse. If a second pulse occurs during the transmission, it is not recognized.

7.3 XON / XOFF CONTROL

Applicable to RS232, not RS485. A measurement transmission is enabled by the receipt of an

ASCII XON character. It is disabled by the receipt of an ASCII XOFF character.

7

. COMMAND MODE

8.1 OVERVIEW

In the Command Mode, the device does not send any data automatically, but responds to com-

mands received from a host computer. These commands can be:

� To transmit the latest or peak measurement
� To reset the meter completely or just the peak value and/or latched alarms
� To display a value sent from the computer
� To transmit present setup parameters
� To receive new setup parameters
� To monitor or alter data in selected memory locations of the meter.

The selection of either the Continuous mode or the Command Mode can be made from the

front panel Menu selection “Ser 2”. The meter will not respond to a command in the Conti-

nuous Mode, except the command “A1”, which puts the meter into the Command Mode.

8.2 COMMAND MODE FORMAT

The minimum format is 4 characters. Example: *5A1
After any command that causes a Meter Reset, such as C0, F, W, X, the Counter sends an “R”

character after the Reset is complete and the Counter is ready to accept a new command.

CHAR 1 - COMMAND IDENTIFIER
All commands begin with “*” followed by the meter address, then a command letter followed

by a sub-command number or letter. Additional characters may be appended. All commands

terminate with <CR> (<LF> ignored). The counter may be assigned a different recognition

character via the RS232 / 485 serial port, but will still recognize the “*”.

Char #

Character

Description

1

2

3

4

*

0-V

A-Z

0-U

Command Identifier (Recognition Character)

Device Address (0 addresses all devices, 1-V specific)

Command Function

Sub-command (or # Bytes or Words of data being transferred)

CHAR 2 - ADDRESS CODES
The next table is the Serial Communication Address Codes following the “*” for each meter

address number. Also shown is the corresponding character that is set in menu item “SER 2”.

15