FUTEK IPM500 (D500) Digital Display User Manual

Page 28

Advertising
background image

1 High adaptive threshold

0 Display batch

1 Display filtered signal

0 Take peak of unfiltered signal

1 Take peak of filtered signal

0 Alarm from unfiltered signal

1 Alarm from filtered signal

35 (U)

Modbus Addr. 00 to FF Modbus address 0-255

(in Hex format)

09 (U)

Setup

Bit 7 6 5 4 3 2 1 0 EXT IN A EXT IN B
0 0 0 Reset Meter Hold

0 0 1 Reset Peak Display

* Both ExtinA & * 0 1 0 Meter Hold Peak Display

ExtinB = * 0 1 1 Meter Hold Tare
Function reset * 1 0 0 Peak Display Tare

1 0 1 Tare Reset

1 1 0 External Decimal Point 1

1 1 1 External Decimal Point 2

1 Coordinates of 2 points for Scale, Offset

0 Scale and Offset direct parameters

0 Normal Format, Ser Com Continuous mode

1 Special Start & Stop characters

0 60 Hz power

1 50 Hz power

0 0 4-1/2 digit display (0.1° for temp.)

0 1 Remote display

1 0 4-1/2 digits count by 10 (0.01° for RTD)

1 1 3-1/2 digit display (1° for temp.)

0D (U)

Alarm Confg4

Bit 7 6 5 4 3 2 1 0 Alarm Trigger Delay

60 Hz 50Hz

0 0 0 0.018 s 0.021 s

0 0 1 0.035 0.043

0 1 0 0.07 0.085

0 1 1 0.14 0.17

1 0 0 0.28 0.34

1 0 1 0.56 0.68

1 1 0 1.13 1.36

1 1 1 2.27 2.72

0 0 0 Al3 Band Dev, Al4 Band Dev

0 0 1 Al3 Hysteresis, Al4 Band Dev

0 1 0 Al3 Band Dev, Al4 Hysteresis

0 1 1 Al3 Hysteresis, Al4 Hysteresis

1 0 0 No deviation in menus or calc

0C (U)

Alarm Confg3

Bit 7 6 5 4 3 2 1 0

0 0 0 0 Al3 Hi active, Al4 Hi active

0 0 0 1 Al3 Lo active, Al4 Hi active

0 0 1 0 Al3 Disabled, Al4 Hi active

0 1 0 0 Al3 Hi active, Al4 Lo active

0 1 0 1 Al3 Lo active, Al4 Lo active

0 1 1 0 Al3 disabled, Al4 Lo active

1 0 0 0 Al3 Hi active, Al4 disabled

1 0 0 1 Al3 Lo active, Al4 disabled

1 0 1 0 Al3 disabled, Al4 disabled

0 0 Al3 non-latch, Al4 non-latch

0 1 Al3 latch, Al4 non-latch

1 0 Al3 non-latch, Al4 latch

1 1 Al3 latch, Al4 latch

0 0 Relay3 On when Al3 active, Relay4 On when Al4 active

0 1 Relay3 Off when Al3 active, Relay4 On when Al4 active

1 0 Relay3 On when Al3 active, Relay4 Off when Al4 active
1 1 Relay3 Off when Al3 active, Relay4 Off when Al4 active

0B (U)

Alarm Confg2

Bit 7 6 5 4 3 2 1 0 Alarm Trigger Delay

60 Hz 50Hz

0 0 0 0.018s 0.021s

0 0 1 0.035 0.043

0 1 0 0.07 0.085

0 1 1 0.14 0.17

1 0 0 0.28 0.34

1 0 1 0.56 0.68

1 1 0 1.13 1.36

1 1 1 2.27 2.72

0 0 0 Al1 Band Dev, Al2 Band Dev

0 0 1 Al1 Hysteresis, Al2 Band Dev

0 1 0 Al1 Band Dev, Al2 Hysteresis
0 1 1 Al1 Hysteresis, Al2 Hysteresis

1 0 0 No deviation in menus or calc

0A (U)

Alarm Confg1

Bit 7 6 5 4 3 2 1 0

0 0 0 0 Al1 Hi active, Al2 Hi active

0 0 0 1 Al1 Lo active, Al2 Hi active

0 0 1 0 Al1 Disabled, Al2 Hi active

28