Dismounting the wall plate, Wall mounting instruction – ROTRONIC CF8-W-Disp-AL User Manual

Page 2

Advertising
background image

Dokument
IMA 210

Rev
1

Page
2 (6)

9 Lid locking screw
10 Wall plate
11 Locking screw
12 Drill mark for cable entry bushing
13 Drill mark for cable entry bushing

Dismounting the wall plate

The sensor is delivered with the wall plate mounted. The wall plate has to be
removed before the sensor is mounted onto the wall. Unscrew the screw on the side
of the box. See figure 3.

Wall Mounting Instruction


Normally the PCB should not be removed from the housing. If for some reason the
PCB must be removed it must be handed carefully and protected from electrostatic
discharge.

1)

Electrical cable entry: The box has a factory mounted cable entry bushing in

dimension PG9. Never feed more than one cable through each cable entry bushing,
or else gas might leak through!

2)

Screw the wall plate onto the wall: The wall plate has holes for three screws.

Drill holes for 3,5mm screws and put dowel into them. Dowels and screws 3,5 x
25mm are included in a plastic bag

3)

Attaching the sensor box to the wall plate is done by a snap-in fitting. The wall

plate has three hooks that fit in holes in the sensor box. Fasten the screw on the side
of the box.

4)

The lid can be locked with the screw at the bottom of the sensor box.

If for some reason the PCB must be removed it must be handed carefully
and protected from electrostatic discharge! Normally, removing the PCB is
not required.