ROTRONIC CF8-W-Disp-AL User Manual

Page 4

Advertising
background image

Dokument
IMA 210

Rev
1

Page
4 (6)

No

Connection

Terminal

Function

Electrical Data

Remarks

1


2

Power (+)


Power ground (-)

24 VAC/DC+ (+-20%),

3W

24 VAC/DC-

2W without output load
See note 1!

3




4




5

Out(1)




Out(2)




Unmarked

Analogue Output

1 (+)


Analogue Output

2 (+)



Signal Ground (-)

0-10 VDC or 0-20 mA
2-10 VDC or 4-20 mA

0-10 VDC or 0-20 mA
2-10 VDC or 4-20 mA

Connected to

via

PTC fuse

According to positions of
OUT1 and start point
jumpers.
See note 2!


According to positions of
OUT2 and start point
jumpers.
See note 2!

See note 1!

6
7
8

NO1

COM1

NC1

(OUT3)

Relay1 Normally

Open

Relay1 Common

Relay1 Normally

Closed

Contact free relay

minimum load 1mA/5V

rated load

0,5A/125VAC;

1A/24VDC

Standard
Open < 1.4 % CO

2

Closed > 1.5 % CO

2


The relay output is potential
free

9

10
11

NO2

COM2

NC2

(OUT4)

Relay2 Normally

Open

Relay2 Common

Relay2 Normally

Closed

Contact free relay

minimum load 1mA/5V

rated load

0,5A/125VAC;

1A/24VDC

Standard
Open < 2.9 % CO

2

Closed > 3.0 % CO

2


The relay output is potential
free

Communi-

cation

jumper

Used to set

communication

protocol

Upper

Modbus communication
protocol

Table I. Electrical terminal connections for

CF8-W-Disp-AL

Note 1: The ground terminal is used as negative power supply DC input or AC phase ground

(halfwave rectifier). The signal ground M, protected by a PTC resistor, is the same as power ground
G0 (permitting a ”3-wire” configuration). A single transformer may be used for the entire system.

Note 2: CF8-W-Disp-AL can deliver a voltage or a current loop for Out(1)/Out(2). To change between
voltage and current output mode the hardware jumpers are used. There is one jumper
for Out(1) and
one for Out(2), so that one output can be a voltage output and the other a current output. Both, voltage
output and current output, can have start points 0 % (0-10 VDC or 0-20mA) or 20% (2-10 VDC or 4-
20mA). The same start point is used for both outputs. See the user manual.


~

+

Advertising