Configurable analog outputs

HM 3-wire

Output 1

Can be made to correspond to any parameter

Factory default parameter

Relative humidity or dew / frost point

Factory default scale

As per ordering code

Output 2

Can be made to correspond to any parameter

Factory default parameter

Temperature

Factory default scale

As per ordering code

Output 1 and Output 2

Signal type

0…20 mA
4… 20 mA
0… 1 V
0… 5 V
0… 10 V
(user configurable)

User configurable scaling limits

-999 … +9999 engineering units

Short circuit tolerant

Yes

Maximum external load

500

Ω (current output)

Minimum external load

1000

Ω (voltage output)

0

Ω (current output)

Digital interface

HM 3-wire

Interface type

RS-485

Service connector

HM 3-wire

Interface type

UART

(Universal Asynchronous Receiver Transmitter)

Maximum service cable length

5 m (16.4 ft)

General specifications

HM 3-wire

HC2-S3 Probe material

Polycarbonate

Probe dust filter material

Polyethylene

Housing material

POM

Housing protection grade

IP 65

Physical dimensions

See Models

Weight (with HC2-S3 probe)

258 g (9.1 oz)

Conformity with standards

HM 3-wire

CE / EMC immunity

EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11

Solder type

Lead free (RoHS directive)

Fire protection class

Corresponds to UL94-HB

FDA / GAMP directives

compatible