3 factory default settings, Factory default settings, E-m-hm-v1_10 – ROTRONIC MP400H User Manual

Page 8

Advertising
background image

E-M-HM-V1_10

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroMet MP Humidity Temperature
Meteorological Probe: User Guide

Instruction Manual

Document Type

Page 8 of 19

Document title

© 2009; Rotronic AG E-M-HM-V1_10

configured, enabled or disabled

DEVICE SAFEGUARDS

Functions

Description

► Device write protection

Used to protect the HygroMet with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled

PROCESS PROTECTION AND PROTECTION OF OTHER DEVICES

AirChip 3000 Functions

Description

► Limit humidity output to 100 %RH

This probe function is used to prevent the humidity signal from
exceeding 100 %RH when condensation forms on the sensor.
Can be enabled or disabled

► Out-of-limit value alarm

Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled

Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display

► Bad sensor alarm

This is a built-in probe function. Cannot be disabled

A bad humidity or temperature sensor triggers a digital alarm,
provided that the HygroMet is set to monitor probe alarms

► Fail safe mode

Used to specify a "safe" fixed value for humidity and for
temperature (HygroMet or probe) in the event of:

o

Loss of communication with the probe (HygroMet function)

o

Sensor failure (probe function)


Can be configured, enabled or disabled

4.3 Factory default settings


Configurable Settings

Factory default

Unit system (Metric or English)

As per ordering code

Analog signal type (4…20 mA or other)

As per ordering code

Psychrometric calculation

As per ordering code

Fixed pressure value

1013.25 hPa or 29.92 In Hg

Output 1 parameter, scale and unit

As per ordering code (%RH or other)

Output 2 parameter, scale and unit

Temperature, unit as per ordering code

Communication protocol

RO-ASCII

RS-485 address

0

Device name

Instrument model

Functions

Factory default

Humidity / temperature adjustment

Device write protection

Disabled (HygroMet and probe)

Limit humidity output to 100 %RH

Enabled (probe)