- All devices within a multi-drop should use the same Baud rate. See Baud rate compatibility

requirements.

- Models with Ethernet (TCP/IP) interface: please note that the Baud rate of the HygroLab and

the Baud rate of the internal Ethernet module should be set to the same value.


RS-485 Cable:
Using a symmetrical transmission method in combination with low capacity/ low
attenuation twisted pair cables, allows extremely reliable long distance connections. The use of a
high grade shielded cable avoids cross talk between the transmitted signals and also reduces the
potential of external interference. For the RS-485 cable, we recommend using a cable
Cat. 5e ANSI/ TIA /EIA-568-A-5.

In general the RS-485 cable should be shielded and comply with the following specifications:

-

Cable capacitance

<=300pF/m or 90 pF/ft

- Line

impedance

100

Ω ±15 Ω

- Line

resistance

140

Ω/km or 225 Ω

-

Signal lines

Twisted pair


In addition, we recommend terminating each end of the RS-485 cable with a 240 Ohm resistor.

RS485 Network:

Addr. 1

Addr. n

RS485 device connectors

240

240






Note: all instruments connected to the network should use the same baud rate

Voltage potential issues: The existence of a voltage-potential between instruments that are
interconnected can be a source on concern in large installations, installations with different mains
power supply and in inter-building networking.

As a first measure, the shield of a signal cable should be connected at both ends. In the case of a
data cable, a low-resistance potential equalization cable may also have to be used. This cable
should be run parallel and as near as possible to the data cable, preferably in the same conduit.
The shield of the data cable should under no circumstances be used as equalization cable! The
conductors of the potential equalization cable should ideally be stranded in order to be effective
also in case of high- frequency interference.