c) AwQuick Mode:

Upon selecting this mode, the HygroLab 3 displays the following:

4

Q D w e l l [ m i n ]


Use the ▼ or ▲ arrow to set the dwell time. In the above example, the HygroLab 3 will wait 4
minutes before processing the humidity data with the AwQuick algorithm.

Default value: 4 min (recommended value)

Press the ENTER key to record and confirm this setting. The display now shows the following:

0. 0 1

Q T S [

° C / m i n ]


Use the UP or DOWN arrow to set the definition of stable temperature. In the above example, the
HygroLab will consider temperature to be stable when the rate of change of the temperature signal
is less than 0.01

°C per minute.


Default value: 0.01

°C / min


Press the ENTER key to record and confirm this setting. The AwQuick mode is now active.

When ready to measure, press the ENTER key to access and start the AwQuick mode. Press the
ENTER key again to stop the AwQuick mode at any time and return to the Standard mode.

Notes:

ƒ

When the HygroLab 3 is in the AwQuick mode, pressing the ENTER key automatically starts

the AwQuick mode, unless the MENU key was pressed immediately before pressing the
ENTER key.

ƒ

The ON / OFF key is inoperative in the AwQuick mode. You must stop the AwQuick mode in

order to be able to turn the power of with the red ON/OFF key.


For instructions on using the AwQuick mode, see Water Activity Measurement.