Setra System 267MR User Manual

Page 12

Advertising
background image

The 4-20 mA current version has only 4 switches.

7.0 MODEL 267 & 267MR PERFORMANCE SPECIFICATIONS

Accuracy RSS

*

(at constant temperature.)

±1.0% FS

Non-Linearity, BFSL

±0.98% FS

Hysteresis

0.2% FS

Non-Repeatability

0.1% FS

*RSS of Non-Linearity, Non-Repeatability and Hysteresis.

Thermal Effects

Compensated Range °F(°C)

+40 to +150 (+5 to +65)

Zero/Span Shift %FS/°F(°C)

0.033 (0.06)

Maximum Line Pressure

10 psi

Overpressure

10 psi in positive or negative

direction

Warm-up Shift

±0.1% FS total

Position Effects

(Unit is factory calibrated at 0g effect in the vertical position)

Range

Zero Offset (%FS/G)

0 to .1” WC

2.1

0 to 1” WC

.22

0 to 5” WC

.14

9

TABLE-2 RANGE SWITCHING INSTRUCTIONS FOR PASCAL RANGES

MR5

MR6 MR7 MR8 MR9 SWITCH SWITCH SWITCH

RANGE RANGE

RANGE

RANGE

RANGE

SETTINGS

SETTINGS

SETTINGS

0-5V 0-10V

4-20MA

OUTPUT

OUTPUT

OUTPUT

Factory Default Setting

0-25Pa 0-200Pa 0-1000Pa 0-2500Pa 0-7500Pa

0-100Pa

0-500Pa 0-1250Pa 0-3750Pa

0-50Pa 0-250Pa 0-625Pa 0-1875Pa

±

12.5Pa

±

100Pa

±

500Pa

±

1250Pa

±

3750Pa

±

50Pa

±

250Pa

±

625Pa

±

1875Pa

±

25Pa

±

125Pa

±

312Pa

±

937Pa