9. Maintenance and inspection

Symptom

Check item

Inaccurate measurement

Is the sensor normal? (Connect a different sensor to the line. Does it show a
similar symptom?)
Is the sensor correctly connected?
Is the correct sensor type set? (Is the sensor the same as the type setting
entered in the product?)
Are all the PV correction values set to the correct levels?

The setting does not match
the measurement.

Does the heater have a sufficient capacity?
Is the integral time (I) set to the correct setting?

Poor control

Is the PID value correct? Perform auto-tuning again.

Abnormal output
(control/event output)

Is the output terminal correctly connected?
Is the control type set to the correct setting?

No communication occurs

Are the switches SW1 and SW2 set to the correct unit number and
communication speed?