WARNING

When performing a cauterization or similar treat-

ment, the patient must be insulated from the metal

portions of the table by nonconductive material.

Failure to do so may result in electrical shock or

burns to the patient.

WARNING

Use 120 VAC, 60 Hz alternating current only. Failure

to do so may result in electrical shock to personnel

and will result in damage to the table.

WARNING

Do not use this table in an explosive or oxygen-en-

riched atmosphere. Failure to do so may result in

serious personnel injury or death.

CAUTION

Do not use any power supply other than that

listed on the rating label (Figure 1). Failure to do

so may result in serious injury and/or equipment

damage.

SPECIFICATIONS

Electrical Requirements

... . . . . . . . . .120 VAC nominal, 60 Hz, 3 amps maximum

Fuse rating.................................3 Amp/250V/P/N 100091
Power Cord

................................ Extends 78 in. (Minimum) from table

3 conductor, 120VAC. With hospital grade plug
Type of protection against electrical shock ........... Class 1
Type of protection

against electrical shock ............................ B applied parts
Type of protection against ingress of water……. Ordinary

Can accept paper rolls of.....................21 in. x 3.5 in. dia.
Certifications .............................UL/CUL Classified 60601

Pelvic Tilt ...............................................................5° Rise
Drawer Warmer ..................120 VAC, 65 watts, 0.5 amps

Weight of Table .....................................................190 lbs.

(With carton and skid)...........................................235 lbs.
Table Weight Capacity ..........................................325 lbs.
Stirrup Extension (maximum): ................................. 16 in.
Stirrup Lateral Positions ......................4 Lateral Positions
Leg Extension (maximum) ....................................... 15 in.
Dimensions:
Upholstered Top ....................28.0 in. wide x 57.3 in. long
With Leg Extension fully extended.......................71.0 in.

Seat Height ...........32 in to top of upholstered seat at foot
Step .................. 7.2 in. high x 20.38 in. wide x 11 in. long
Step Capacity .......................................................325 lbs.
Base ............ 28.2 in. high x 23.25 in. wide x 46.0 in. long
Transportation and Storage
Temperature .............................................-28°C to 80°C
Humidity......................................................-10% to 90%
Operation
Temperature .............................................. 10°C to 40°C
Humidity.......................................................30% to 75%

ELECTRICAL REQUIREMENTS

CAUTION

This product has been evaluated with respect to

electrical shock, fire, and mechanical hazards only

in accordance with UL60601-1; IEC60601-1;

CAN/CSA C22.2 No. 601.1

Electromagnetic Interference

This product is designed and built to minimize electromagnetic interference with other devices; however, if interfer-

ence is noticed between another device and this product, remove the interfering device from the room or plug this
product into an isolated circuit.