3 functional specifications – AirSep MN138-1 H User Manual

Page 11

Advertising
background image

5

MN138-1 Rev H

02/14


1.3 Functional Specifications

Oxygen Concentration:

1-5 liters per minute at 90% +5.5/-3%

1-3 liters per minute at 90% +5.5/-3% (VisionAire 3)

-2 liters per minute at 90% +5.5/-3% (VisionAire 2)

(Based on 70°F [21°C] at sea level)

Accuracy:

± 10% of indicated flow setting, or ± 200 ml, whichever is greater.

Response Time:

Allow 5 minutes to attain maximum oxygen concentration.

Positioning:

Operate the unit in an upright position, maintaining at least 12 inches
(30.5cm) of open space on all sides for ventilation.

Dimensions:

20.8 in. high, 14.1 in. wide, 11.5 in. deep

(52.8 cm high, 35.8 cm wide, 29.2 cm deep)

Weight:

30 lb; shipping weight – 37 lb

(13.6 kg; shipping weight – 16.7 kg)

Electrical

(+/- 10%)

:

115VAC, 60 Hz, 3 amps, 290 watts

115VAC, 60 Hz, 3 amps, 190 watts

(VisionAire 3)

115VAC, 60 Hz, 3 amps, 200 watts

(VisionAire 2)

220-240 VAC, 50 Hz, 1.5 amps, 290 watts

220 VAC, 60 Hz, 1.5 amps, 290 watts

230 VAC, 50 Hz, 1.5 amps, 180 watts (VisionAire 2)

230 VAC, 50 Hz, 1.5 amps, 175 watts (VisionAire 3)

Alarms:

Power Failure

Low Oxygen Concentration (with optional Oxygen Monitor)
Low Pressure
High Pressure

High Temperature

No Flow

Electrical Shock

Class II, Type B

Protection:

Operating Temperature

41 to 105°F (5 to 40°C)

Range:

Storage Temperature

-4 to 140°F (-20 to 60°C)

Range:

Humidity:

15 – 95% (non-condensing)



Advertising