AirSep MN138-1 H User Manual

Page 34

Advertising
background image

28

MN138-1 Rev H

02/14

5.3 Troubleshooting Chart

Problem

Probable Cause

Solution


Unit does not run.
I/0 power switch in “I” position.

No power to unit.

Unit circuit breaker tripped/faulty

Faulty electrical connections.

Faulty circuit board.

Faulty I/0 power switch.

Check wall outlet for power.

Reset or replace circuit breaker.

Check electrical connections.

Replace circuit board.

Replace I/0 power switch.


Compressor runs with intermittent
alarm. Oxygen concentration at max
flow rate is within specifications.



• Faulty pressure transducer.



• Replace circuit board.

Compressor runs and shuts down
periodically. Only start up alarm
activates


• Faulty oxygen monitor circuit board.


• Replace oxygen monitor circuit board.


Compressor runs and shuts down
periodically with 5-beep intermittent
high temperature alarm.

Restricted air flow.

Unit overheating due to improper
location. Compressor thermally cut off
due to excessive heat. NOTE: It will
not restart until it cools down.

Faulty cabinet fan.

Temperature probe wire harness faulty
or disconnected from circuit board.

Remove obstruction.

Locate unit away from heating source,
providing adequate ventilation on all
sides.

Do not operate more than one unit
side by side.

Replace cabinet fan.

Replace or reconnect temperature
probe.


Compressor runs with 3-beep
intermittent low pressure alarm
and low oxygen concentration.

 

Leak

 

Reduced air intake

Weak compressor

Faulty circuit board.

Leak test and repair leak.

Check compressor intake path for
obstruction. Clean/remove
obstruction.

Replace compressor.

Replace circuit board

.



Compressor does not start. I/0
power switch in “I” position,
intermittent 3-beep low pressure
alarm, and cabinet fan turns.



Extreme cold start.

Faulty electrical connection for
compressor.

Faulty capacitor.

Faulty circuit board.


Allow unit to reach room
temperature.

Check electrical connections for
compressor.

Replace capacitor.

Replace circuit board

4-beep high pressure alarm is
activated and both compressor
and valves do not operate.


Restriction in exhaust muffler.

Contaminated sieve beds.

Faulty solenoid valve.

Faulty circuit board.


Replace exhaust muffler.

Replace sieve beds.

Replace solenoid valve manifold.

Replace circuit board.