Boonton 4540 Series RF Power Meter

Remote Operation

5-40

OUTPut:EXTernal:PULse:PERWID

Description:

Set or return the pulse period <arg1> and pulse width <arg2> in seconds in the Variable pulse
control mode of the external 1 GHz calibrator. This parameter has no effect in the preset pulse
control mode or the external pulse source mode.

Syntax:

OUTPut:EXTernal:PULse:PERWID <arg1> <arg2>

Argument:

<arg1> = 28e-6 to 131.07e-3 (Pulse period)

<arg2> = 7e-6 to 65.535e-3 (Pulse width)

Valid Modes:

Any

OUTPut:EXTernal:PULse:SOURce

Description:

Set or return the pulse modulation source for the external 1 GHz calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator’s rear-panel EXT
PULSE
input. Note this input requires TTL compatible signal levels.

Syntax:

OUTPut:EXTernal:PULse:SOURce <character data>

Argument:

<character data> = {INT, EXT}

Valid Modes:

Any

OUTPut:EXTernal:SIGNal

Description:

Set or return the on/off state of the external 1 GHz calibrator output signal.

Syntax:

OUTPut:EXTernal:SIGNal <Boolean>

Argument:

<Boolean> = { 0, 1, OFF, ON }

Valid Modes:

Any

OUTPut:INTernal:LEVel[:POWer]

Description:

Set or return the power level of the internal 50 MHz calibrator output signal.

Syntax:

OUTPut:INTernal:LEVel:POWer < numeric_value >

Argument:

<character data> = -60.0 to +20.0 dBm (0.1dB resolution)

Valid Modes:

Any