Boonton 4540 Series RF Power Meter

Operation

4-26

Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle

Description:

Sets the duty cycle in percent for calculated CW pulse power measurements. Valid only for

thermal sensors and CW sensors in the square-law region and subject to the accuracy of the duty

cycle value. Setting the duty cycle to 100% is equivalent to a CW measurement. Note that this

method of measuring pulse power should be used only if a peak power measurement cannot be

used. This item is not pictured above—it only appears in Modulated mode. The Duty Cycle

Menu item is disabled for Peak Power Sensors.

Menu Type:

Numeric Value

Range:

0.01% to 100.00%

SCPI Command:

SENSe[1|2]:CORRection:DCYCle <numeric_value>





















Fig. 4-19 Corrections

Menu (Pulse Mode)

Voltage Probe














Fig. 4-20 Corrections

Menu (Pulse Mode) Peak

Power Sensor - No TC















Fig. 4-21 Corrections

Menu (Pulse Mode)

CW Power Sensor














Advertising